Product overview
Alternating front and rear mounting supports parallel pipetting at a minimum 18 mm pitch.
- Minimum 18 mm pitch for multi-channel layouts.
- Alternating front and rear mounting for parallel layouts.
- Capacitive and pressure-based liquid-level detection.
- Tip status monitoring throughout the pipetting cycle.
- Air-path blockage detection.
- Active ejection of disposable tips.
- RS-485 and CAN communication.
- Compact command set and host control software.
- Integrated ceramic plunger.
- Low-noise operation.
- Disposable tips help reduce sample carryover.
- Can be integrated with a compatible Z-axis module.
- Bubble detection.
Technical specifications
- Maximum pipetting volume
- 1,000 µL
- Pipetting speed
- 1,000 µL/s
- Power supply
- 24 V DC, 1 A peak
- Electrical connection
- 12-pin FFC
- Communication
- RS-485 / CAN
- Signal output
- One open-collector output for liquid-level detection
- Operating environment
- 15–35 °C, 20–85% relative humidity, non-condensing
- Drive mechanism
- Stepper motor and lead screw drive the plunger, with encoder feedback for closed-loop control
- Plunger travel (total / effective)
- 64 mm / 57 mm
- Dimensions (H × W × D)
- 199 × 56 × 21 mm
- Mass
- 234 g
- Rated service life
- 1,000,000 cycles
- Compatible disposable tips
- 10, 50, 200 and 1,000 µL, conductive or nonconductive
- Minimum multi-channel pitch
- 18 mm
Pipetting performance
| Operating mode | Tip capacity | Test volume | Precision (CV) | Accuracy |
|---|---|---|---|---|
| Single aspiration / single dispense | 10 µL | 1 µL | ≤5% | ±5% |
| Single aspiration / single dispense | 50 µL | 5 µL | ≤1.8% | ±1.5% |
| Single aspiration / single dispense | 200 µL | 10 µL | ≤2% | ±2.5% |
| Single aspiration / single dispense | 1,000 µL | 100 µL | ≤1% | ±2% |
| Single aspiration / multiple dispenses | 1,000 µL | 20 µL | ≤2.5% | ±3% |
CV is the coefficient of variation. Results are shown for the stated test volume and tip capacity where available.
Dimensions