Product overview
Large-volume air-displacement pump with pressure-based liquid-level detection and a precision syringe.
- Large-volume operation over a broad liquid-transfer range.
- Pressure-based liquid-level detection.
- Tip status monitoring throughout the pipetting cycle.
- Air-path blockage detection.
- Active ejection of disposable tips.
- Detects aspiration of air.
- RS-485 and CAN communication.
- Compact command set and host control software.
- Integrated precision syringe.
- Low-noise operation.
- Disposable tips help reduce sample carryover.
- Can be integrated with a compatible Z-axis module.
Technical specifications
- Maximum pipetting volume
- 10,000 µL
- Pipetting speed
- 3,000 µL/s
- Power supply
- 24 V DC, 1 A peak
- Electrical connection
- 7-core high-flex cable
- Communication
- RS-485 / CAN
- Signal output
- One open-collector output for liquid-level detection
- Operating environment
- 15–35 °C, 20–85% relative humidity, non-condensing
- Drive mechanism
- Stepper motor and lead screw drive the plunger, with encoder feedback for closed-loop control
- Plunger travel (total / effective)
- 72 mm / 64 mm
- Dimensions (H × W × D)
- 299.5 × 71.5 × 31 mm
- Mass
- 1,000 g
- Rated service life
- 1,000,000 cycles
Pipetting performance
| Operating mode | Test volume | Precision (CV) | Accuracy |
|---|---|---|---|
| Single aspiration / single dispense | 20 µL | ≤4% | ±8% |
| Single aspiration / single dispense | 50 µL | ≤2.5% | ±5% |
| Single aspiration / single dispense | 500 µL | ≤1.5% | ±2% |
| Single aspiration / single dispense | 5,000 µL | ≤1% | ±1.5% |
| Single aspiration / single dispense | 10,000 µL | ≤0.5% | ±1% |
| Single aspiration / multiple dispenses | 100 µL | ≤3% | ±5% |
| Single aspiration / multiple dispenses | 500 µL | ≤2.5% | ±2% |
CV is the coefficient of variation. Results are shown for the stated test volume and tip capacity where available.
Dimensions