Product overview
A compact pump for 5 mL disposable tips. Parallel layouts support a minimum 18 mm pitch.
- Capacitive and pressure-based liquid-level detection.
- Tip status monitoring throughout the pipetting cycle.
- Air-path blockage detection.
- Active ejection of disposable tips.
- Detects aspiration of air.
- Low-noise operation.
- Disposable tips help reduce sample carryover.
Technical specifications
- Maximum pipetting volume
- 5,000 µL
- Pipetting speed
- 5,000 µL/s
- Power supply
- 24 V DC, 1 A peak
- Electrical connection
- FFC. Confirm pin count and pinout before integration
- Communication
- RS-485 / CAN
- Signal output
- One open-collector output for liquid-level detection
- Operating environment
- 15–35 °C, 20–85% relative humidity, non-condensing
- Drive mechanism
- Stepper motor and lead screw drive the plunger, with encoder feedback for closed-loop control
- Plunger travel (total / effective)
- 43.5 mm / 38 mm
- Dimensions (H × W × D)
- 197 × 72 × 18 mm
- Mass
- 426 g
- Rated service life
- 1,000,000 cycles
- Minimum multi-channel pitch
- 18 mm
Pipetting performance
| Operating mode | Tip capacity | Test volume | Precision (CV) | Accuracy |
|---|---|---|---|---|
| Single aspiration / single dispense | 5 mL | 10 µL | ≤3% | ±5% |
| Single aspiration / single dispense | 5 mL | 50 µL | ≤2.5% | ±4% |
| Single aspiration / single dispense | 5 mL | 500 µL | ≤1% | ±2% |
| Single aspiration / single dispense | 5 mL | 1,000 µL | ≤1% | ±1.5% |
| Single aspiration / single dispense | 5 mL | 3,000 µL | ≤1% | ±1% |
| Single aspiration / single dispense | 5 mL | 5,000 µL | ≤0.5% | ±1% |
| Single aspiration / multiple dispenses | 5 mL | 50 µL | ≤3.5% | ±5% |
| Single aspiration / multiple dispenses | 5 mL | 500 µL | ≤2% | ±2% |
CV is the coefficient of variation. Results are shown for the stated test volume and tip capacity where available.
Dimensions